HVIT Calibration System

 HVIT calibration system

 

The above pictured HVIT Calibration System is made up of 1 Philips-Fluke

Frequency Counter, 1 Speed-Trap Light Source/Sensor and 1 Custom Electronic

Interface/Power Source. The Speed-Trap is mounted on the inside panel of the

existing HVIT (at the end of the barrel) and test shots are made comparing the

HVIT Velocity Readout and the speed in micro-seconds of the Frequency

Counter. The micro-second reading is converted to velocity and the two readings

are compared. If the readings are significantly different, adjustments are made on

the HVIT to correct it's readings tp match those on the Frequency Counter. After

calibration is completed a series of 20 shots are fired to establish a qualified

Satistical Data comparison.

 

This unit can be rented or purchased.